EasyManua.ls Logo

Vernacare Vortex - Fault 10: Machine Not Clean Internally

Vernacare Vortex
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Product stock
has large
particle size
over 10 mm
Lack of water in the
hopper
Check isolation stop valve is open.
Check water pump is operating
See fault 3.
Check non-return valve is sealing
Replace.
Check float valve is not blocked Clean.
Check water filters are not blocked
Clean.
Blockage in lid spray
Check spray washer for blockage Clean.
Check lid arm water inlet block for
blockage Clean.
Check connection of pipe to lid arm for
blockage - clean
Blockage in water feed
pipe work.
Check pipes from water pump for
blockage Clean.
Check water pump feed pipe for
blockage Clean.
Check connection of pipe to lid arm
for blockage Clean.
Diaphragm outlet
failure.
Check diaphragm condition Replace.
Check solenoid water valve is operating
Replace.
Check water solenoid valve filters for
blockage Clean.
Check diaphragm valve spring is not
damaged or detached Replace.
Blocked drain.
`BLOCKED DRAIN`
Drain is blocked Clean.
Service drain is blocked Clean.
Blade is worn.
Check blade edges for damage
Replace.
3 Phase motor rotating
in wrong direction.
Check direction of cutter blade rotation
Clockwise when viewed from above
Reverse wiring as necessary.
Fault 10