13
4. Immediately touch the tip of the test strip to the drop of blood. The blood pulled
into the test strip through the tip. Make sure that the blood sample has been
fully filled the check window on the strip tip. Hold the tip of the test strip in the
blood drop until the meter beeps.
5. Immediately touch the tip of the test strip to the drop of blood. The blood pulled
into the test strip through the tip. Make sure that the blood sample has been
fully filled the check window on the strip tip. Hold the tip of the test strip in the
blood drop until the meter beeps.
Note:
If the blood sample does not fill the check window, do not add a second drop.
Discard the test strip and start over with a new test strip.
6. The meter counts down 5 seconds and your result appears on the display after
a beep. The test result will automatically be stored in the meter memory.
Please do not touch the test strip during the countdown as this may result in an
error.
Discard the Used Test Strip
You can discard the used test strip by hand. The Meter turns off automatically after
a beep.