14. Technical data
Display ..............................................................6000 counts (characters)
Measuring rate ..................................................approx. 3 measuring operations/second
Measuring method AC ......................................True RMS, AC-coupled
Measuring line length .......................................each approx. 80 cm
Measuring impedance ......................................≥10MΩ(V-range)
Measuring jacks distance .................................19 mm (COM-V)
Automatic deactivation .....................................approx. 15 minutes, can be deactivated manually
Voltage supply ..................................................9 V block battery (NEDA 1604, 6F22 or same build)
Operating time/battery ......................................approx. 72 h (without display lighting/LED lamp)
Working conditions ...........................................0 °C to +30 °C (<75%rH)
..........................................................................+31 °C to +40 °C (<50%rH)
Operating height ...............................................max. 2000 m
Storage temperature .........................................-10 °C to +50 °C (<75%rH)
Weight ..............................................................approx. 473 g
Dimensions (LxWxH) ........................................195 x 95 x 58 (mm)
Measuring category ..........................................CAT III 1000 V; CAT IV 600 V
Degree of contamination ..................................2
Safety according to ...........................................EN61010-1
Protection type ..................................................IP65 (dust-tight and jet-water-protected)
Measurement tolerances
Statement of accuracy in ± (% of reading + display error in counts (= number of smallest points)). The accuracy is
valid for one year at a temperature of +23 °C (±5 °C), and at a relative humidity of less than 75%, non-condensing. A
temperaturecoefcientappliesoutsideofthistemperaturerange:+0.1x(speciedaccuracy)/1°C.
Themeasurementcanbeimpairedwhenthedeviceisoperatedwithinahigh-frequencyelectromagneticeldstrength.
In an electromagnetically burdened environment up to 1 V/m, the accuracy increases by 5% of the measured value.
Theareaabove1V/misnolongerspeciedandmaycausewrongdisplays.