5.4 High voltage tank
- Tube voltage: 100 kV constant potential
- Tube current: 1 mA direct current
5.5 X-ray exposure area
- Type: Square
- Source to Image receptor Distance (SID): 650 mm
- X-ray field size: 430 mm X 430 mm at SID 650 mm
5.6 Re-chargeable battery
- Model name: FB703562-061PBTC
- Type: Li-ion polymer Battery
- Rating: 22.2 Vd.c (Charging voltage 25.2 Vd.c)
- Capacity: 1,460 mAh
- Size: 66(Length)42(Height)39(Width) mm
3
5.7 AC/DC adapter Specification
- AC/DC adapter provided by VSI is a part of VX-30.
- Model name: FY2550900
- Rated input: 100-240 Va.c., 50/60 Hz, 0.4 A
- Rated output: 25.5 Vd.c., 0.9 A
5.8 Software for VX-100
- Type: Built-in
- S/W name: ClaroxV100ControlFirmware
- S/W version: V1.0.0
5.9 Extra accessory
- Cradle
- Wrist strap
5.10 Minimum requirement for digital X-ray image receptor
- Min. resolution: more than 1000
- Min. size: more than 430 mm × 430 mm
- Max. pixel pitch: less than 300 μm
5.11 Protection against Residual Radiation
- To avoid residual radiation caused by using of VX-100, the operator should stay in the Significant
zone of occupancy described in section 4.2 of this user manual and the alinement between the patient
and VX-100 should be kept like [Figure 6].
5.12 Metrics about imaging performance
- To keep the imaging performance, The following parameters should be measured once for every
year and performed by an authorized person or manufacturer.
1) Tube voltage: measurement point 70-100 kV / Tolerance ± 10 %
2) Tube current: measurement point 1 mA / Tolerance ± 20 %
3) Exposure time: measurement point 0.10 s, 1.0 s / Tolerance ± (10 % + 1 ms)
4) Leakage radiation