6. Maintenance
6.1 Replacement of Rechargeable battery
- Unfasten bolts(See the [Figure 8] above.) from the battery cover located on the bottom of the device.
- Take out the battery from the main unit.
- Disconnect the battery connector and change the new battery.
Use only specified battery provided by manufacturer.
The replacement should be performed by authorized person only.
The batter
should be performed periodic checkin
or replaced.
6.2 Periodic inspection (Quality Control Procedure)
We recommend to check this equipment annually.
Only qualified people can check this equipment.
Check items according to the Regulations of the country.
- Inspection period: 1 time / 1 year
- If the result is not satisfied the criteria, please contact to manufacturer.
Inspection
item
Method Criteria
Tube voltage
Place the voltage measuring device at (25 ± 2) cm
away from the focus point, set the device to 70-100
kV, and measure the value of irradiating the X-ray.
Within 70 - 100 kV ± 10 %
Tube current
Exposure time
Open the battery cover. Connect the oscilloscope to
current measurement terminal. Set the device to 1
mA(= 4 V), and measure the value of irradiating the X-
ray.
Within 4 V ± 20 %
Within 0.10 - 1.0 s
± (10 % + 1 ms)
Battery voltage
Open the battery cover. Connect the oscilloscope to
battery terminal and measure the value of battery DC
voltage.
More than 20 Vdc