56 EX1200 Series Preparation for Use
Restore the EX1200’s Default Network Settings
It may be the case that the EX1200 is in an unknown network configuration. The EX1200 can be
returned to its default state by pressing the Reset Button located at the rear of the chassis.
1) Power off the EX1200.
2) Press and hold the reset button until the LAN LED flashes green (approximately 5 s).
3) Release the reset button.
Determine PCs Network Settings
1) Ensure that all host PC Ethernet connections are made that will be used while controlling the
EX1200. If the PC will be connected to a LAN, ensure that the LAN connection is made.
2) From Windows , navigate to StartSettingsNetwork Connections.
3) Right click on the connection that will be used to communicate with the EX1200, then select
Properties (see Figure 2-14). Some PCs have multiple network connections, so it is important
to make certain that the connection used to communicate with the EX1200 is selected.
FIGURE 2-14: NETWORK CONNECTION PROPERTIES
4) Select Internet Protocol (TCP/IP), then click the Properties button as shown in Figure 2-15.
FIGURE 2-15: LOCAL AREA CONNECTION PROPERTIES DIALOG BOX