OPERATING INSTRUCTIONS
The Incubating/Cooling Shaker has been designed to heat/cool and shake
PCR plates in the standard or manual mode.
The temperature, speed and time functions work independently of one
another. The temperature and speed can be re-set without re-setting the
timer and the timer can be stopped and started without interrupting the
heating/cooling and shaking functions.
MODE SELECTION
This unit can run in three modes; Standard/Manual mode, User
Programmable mode, or a predefined Profile mode. The unit will power
up in whatever mode was last selected.
To change modes: With unit in Standby, hold Select Down Arrow (I) and
Abort (M) buttons, then press the Standby (A) button. Use the “Select”
button to display the desired mode. STD is the standard or manual mode
that allows the user to select settings for temperature, speed, and or
time. Pr1 is a predefined profile. Pr2, Pr3, Pr4, and Pr5 are 4 programs
available to be user defined. These programs must be defined and saved
prior to being avaiable as a selection. The time display will show “USEd”.
When the desired program is displayed, press the Start (H) button to
begin operation.
See page 8 for user defined programming instructions and page 13 for
running the predefined program.
STANDARD/MANUAL MODE
1. Getting ready:
a. Plug the cord into a properly grounded, 3-pronged outlet. The standby
indicator light will illuminate, and a single audible beep will sound,
SPECIFICATIONS
Overall dimensions: 17.9 x 11 x 10.5"
(L x W x H) (45.5 x 27.9 x 26.7cm)
Electrical(50/60 Hz): 120 volts, 2 amps, 200 watts
230 volts, 1 amp, 200 watts
Fuses: 5mm x 20mm, 5 amp quick acting
Temperature range:
10°C below ambient to 95°C ,when
operating conditions are 15º to 40ºC
(Standard and Program Mode
)
45°C and 68°C (Profile Mode)
Temperature uniformity: ±0.5°C at 37°C
Speed range running: 100 to 1600rpm
Speed accuracy: ±2% of set speed up to 999rpm
±5% 1000 to 1600rpm
Orbit: 3mm
Capacity: 2 PCR plates
Timer: digital, 1 minute to 9999 minutes
(increased in 1 minute increments)
Controls: see page 4
Ship weight: 34lbs (15.4kg)