The CP 2M’s menu consists of seven main headings, corresponding to the analyzer’s most
important functions (illus. 7).
When switching on the CP 2M, be sure to allow the software to fully load before taking dew
point measurements.
The software must also be fully loaded before making any menu selections.
«measure: HC»
In this mode, the CP 2M measures the value of dew point of hydrocarbons (dpHC).
Confirm this selection by touching the «►» button with the magnetic key.
To end this selection, touch the «◄» button.
«measure: W»
In this mode, the CP 2M measures the dew point of water (dpW).
Confirm this selection by touching the «►» button.
To end this selection, touch the «◄» button.
«measure: W & HC»
In this mode, the CP2M alternately determines the dpHC and the dpW.
Confirm this selection by touching the «►» button.
To end this selection, touch the «◄» button.
«calibr: HC»
This activates the calibration mode of the hydrocarbon measurement channel.
Using this function, the operational sensitivity of the dpHC measurement channel can be
adjusted as desired.