Cong Prima 2M - manual
33
!
Should there still be residual contamination after manual cleaning has been completed, the
cleaning procedure, as described above, can be repeated using acetone in place of the
cleaning fluid delivered with the analyzer.
5.2 Cong Prima 2M technical information
Dew point measurement range
dpW
T
(Housing)
-40 °C
dpHC
T
(Housing)
-40 °C
Absolute error limit of dew point measurements
dpW
± 0.5 °C
dpHC
± 0.5 °C
Measurement cycle duration period
dpW
10 30 min
dpHC
10 30 min
Gas sampling conditions:
Pressure
160 bar
Temperature
20 +50 °С
Technical data of the CP 2M:
Electrical connection
Conductors: 4 х 0.75 mm
2
(ø 8 11.5 mm)
Material in contact with
the sample gas
Stainless steel 12Х18Н10Т;
PTFE 4;
Glass;
Silicone
Average sample-gas flow volume
0.3 0.5 Nl/min
1)
Explosion protection
IIG EEx d IIA T5.
Warning!
Please use the utmost caution when cleaning the
condensation mirror manually.
Damage such as scratches or etching significantly
reduce the analyzer’s performance.
Mechanical damage to the condensation mirror is
not covered under the warrantee.