31
!
5)
Malfunction in the CP 2M’s
mirror temperature measurement
channel
Defective temperature
sensor
The CP 2M must be
returned to the
manufacturer for
servicing.
6)
Malfunction of the
thermoelectric battery (TEB)
There is no effective
heating or cooling of the
condensation mirror
The CP 2M must be
returned to the
manufacturer for
servicing.
7)
Sensor contamination
The level of the data signal
for water or hydrocarbons
exceeds the allowable
range
Clean the CP 2M’s
mirror.
8)
Malfunction of the laser diode
The temperature of the
laser diode is below the
allowable range
If the housing
temperature (T
B
) is
- 20°C return the
analyzer to the
manufacturer for
servicing.
9)
Contamination still present
after the cleaning mode cycle
is complete
No evaporation of
condensate residue after
three successive cycles of
the cleaning mode have
been completed
Turn off the analyzer
for at least one hour.
Should the error
message be
displayed at the
conclusion of this
period, the analyzer
must be returned to
the manufacturer for
servicing.
If any of the listed error messages appear, it is recommended that the corresponding
remedial action be taken.
If, after taking the appropriate remedial action, the Cong Prima 2M does not function properly,
it will be necessary to return the analyzer to the manufacturer for servicing.