Configuring the e-SAT/IN module 3-9
Setting units
Use the Units field to select the type of unit to appear on the y-axis of the
chromatogram (for example, mV or AU). Units multiplied by Scale Factor
equals 1 V. The maximum Units label length is six alphanumeric characters.
The default is mV.
Setting the Scale Factor
Use the Scale Factor field to select the value that Empower software uses to
adjust the display of the signal along the y-axis of the chromatogram. This
value associates the Units parameter to volts (V). Scaling Factor multiplied by
Units equals 1 V. For example, if you require AU on the y-axis and are using a
detector where 1 AU equals 1 V, enter a scale factor of 1.0. The default setting
is 1000.
e-SAT/IN Events tab
Use the Events tab to define operation and initiation of the selected e-SAT/IN
module output events.