Average interval: time interval in seconds for calculating an average. Value
between 1 and 1800 seconds (60 minutes).
Note: between 6 and 30 minutes the allowed interval is in multiples of 5 seconds.
Between 30 minutes and 60 minutes, in multiples of 10 seconds.
Enable snooze alarm: with this option activated, when the ON/OFF button is
pressed after an alarm sounds, it will be muted for 1 minute.
Enable vibrate on double tap function: if this option is activated and your
WaveMon RF is on, the device will vibrate every time the user pokes it twice.
Second section: Program parameters:
Date format: used to select the day/month/
year format.
Time format: 12 or 24 hours.
Language: used to select the language.
The program must be rebooted to apply the
changes made.
Third section: Updating the rmware for the device:
Select the .bin le of the new rmware and wait
for the device to be updated. Do not switch the
device off while the rmware is being updated
and make sure that the device has sufcient
battery charge before starting the update.
Fourth section: Limit:
Some models have the option to select the
limit for public or workers. Please note that
public limits are not compatible with the
magnetic sensor so only electric eld will be
registered if a public limit is selected.