12. MODELS
12.1 WaveMon RF-8
Part number Version Description
WWM0001 WaveMon RF-8 ICN
*
300 kHz - 8 GHz ICNIRP 1998/2020
*
Occupational &
General public
WWM0002 WaveMon RF-8 FCC 3 MHz - 8 GHz FCC Occupational & General public
WWM0003 WaveMon RF-8 SC6 10 MHz - 8 GHz SC6 2015 Occupational
WWM0004 WaveMon RF-8 EUD 300 kHz - 8 GHz EMF Directive 2013/35/EU &
EU recommendation 1999/519/EC
*
Compliant with ICNIRP 2020 from 27 MHz.
E eld H eld
WaveMon Version
Response shaped to
Frequency range & response Frequency range & response
WaveMon RF-8 ICN
*
ICNIRP 1998/2020
*
±3.5 dB (300 kHz - 8 GHz) 25 MHz - 1 GHz
±1 dB (30 MHz - 700 MHz)
+0 dB/-2.5 dB (700 MHz - 850 MHz)
+0 dB/-5 dB (850 MHz - 1 GHz)
WaveMon RF-8 EUD
EMF Directive
2013/35/EU & EU
recommendation
1999/519/EC
WaveMon RF-8 FCC
FCC 96-326
±3.5 dB (3 MHz - 8 GHz) 3 MHz - 1 GHz
±1 dB (3 MHz - 400 MHz)
±2 dB (400 MHz - 850 MHz)
+0 dB/-4 dB (500 MHz - 1 GHz)
WaveMon RF-8 SC6
Safety Code 6
±3.5 dB (10 MHz - 8 GHz) 25 MHz - 1GHz
±2 dB (30 MHz - 800 MHz)
+0 dB/-5 dB (800 MHz - 1 GHz)
*
Compliant with ICNIRP 2020 from 27 MHz.
E eld H eld
Dinamic range 1 - 300% 2 - 300%
Linearity 0.5 dB (2% – 200%) ±1 dB (5% – 200%)
Isotropic deviation ±1 dB @ 1 GHz ±1 dB @ 400 MHz
Note: the H eld sensor is used only for the Occupational limits, as only workers can
be found in near eld condition where both E and H eld are measured.