18-6 Model 9100 — Operator's Guide: Error Messages
Error Messages
18
F
Failed -------- D
Failed to clear Flash RAM A
Failed to save (cjc) factor D
Failed to save (R-eqV) factor D
Failed to save configuration F
Flash RAM protected by switch A
Frequency outside limits D
Frequency too big B
Frequency too small B
Freq too big: 1kHz max when delta Φ active B
Freq too big: 1kHz max when Φ ref out active B
G
Gain request limited
Requested gain setting is beyond the maximum
capability of the DAC. The maximum value has
been set.
D
H
High volt lvl below low volt lvl
The voltage specified for the high level of a pulse
waveform is below that specified for its low level
B
High volt lvl equals low volt lvl
The voltage specified for the high level of a pulse
waveform is the same as that specified for its
low level
B
Hours entry is not a valid number F
I
Illegal duty/width combination B
Illegal offset B
Illegal rep/width combination B
Incorrect or missing workmat.
Cold junction default = 25°CE
Invalid CJC measurement - default to 25°CE
Invalid junction measurement E