WM 67841c 03/2015
WM 100 TD EN 87
12 Appendix
prisma25S
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Available modes
4 hPa to 25 hPa
4 hPa to 25 hPa
20% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
CPAP, APAP, S, autoS
prisma25ST
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Backup frequency
- Available modes
4 hPa to 25 hPa
4 hPa to 25 hPa
20% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
Auto, 0 bpm to 35 bpm
CPAP, APAP, S, autoS, autoS/T, S/T, T
prisma30ST
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Ti
- Trigger inspiration
- Trigger expiration
- Pressure rise rate
- Pressure drop rate
- Backup frequency
- Target volume
- Pressure adjustment
- Available modes
4 hPa to 30 hPa
4 hPa to 25 hPa
20% to 67%
500 ms to 4000 ms
Auto, can be set to 3 levels
Auto, can be set to 3 levels
Can be set to 4 levels
Can be set to 3 levels
Auto, 0 bpm to 35 bpm
300 ml to 2000 ml
Can be set to 3 levels
CPAP, APAP, autoS/T, S, S/T, T, aPCV
prisma30ST-C
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Ti
- Trigger inspiration
- Trigger expiration
- Pressure rise rate
- Backup frequency
- Available modes
4 hPa to 30 hPa
4 hPa to 25 hPa
20% to 67%
500 ms to 4000 ms
Auto, can be set to 3 levels
Auto, can be set to 3 levels
Can be set to 4 levels
0 bpm to 35 bpm
CPAP, S, S/T, T, aPCV
Specification Therapy device