Note One gram (0.001 kg) is equal to 1 cc of liquid. (1.0 specific
gravity).
To ensure the most accurate measurements, do the following:
Perform the procedure each time a urine output measurement is needed.
Do not rely on subtracting a known weight for a dry diaper from the displayed weight
of a wet diaper.
Remove any fecal matter from the wet diaper before starting the weighing
procedure. The change in the dry and wet diaper should only reflect urine content.
Send measurements to a computer
The scale can connect to a computer using the Microsoft
®
Windows
®
10 operating
system.
Detailed instructions for this feature are available from Welch Allyn. Go to http://
www.welchallyn.com/en/other/contact-us.html to find your local representative.
CAUTION Any computer connected to the scale must be running on
battery power, a 60601-1-compliant power supply, or a 60601-1-compliant
isolation transformer.
1. Connect the USB Type B male side of the cable to the USB port on the rear of the
scale.
2. Connect the USB Type A male side of the cable into a USB port on your computer.
Windows automatically installs the necessary driver.
3. Press SEND DATA on the scale to send the current measurement to the computer.
Send measurements to a Welch Allyn device
The scale can connect to a Welch Allyn Connex
®
Vital Signs Monitor or a Welch Allyn
Connex
®
Spot Monitor.
CAUTION Any device connected to the scale must be running on battery
power, a 60601-1-compliant power supply, or a 60601-1-compliant isolation
transformer.
1. Connect the USB Type B male side of the cable to the USB port on the rear of the
scale.
2. Connect the USB Type A male side of the cable into a USB port on the rear of your
device.
3. Press SEND DATA on the scale to send the current measurement to the device.