26 Functional verification Welch Allyn VSM 300 Series
6. Verify that the reading displayed on the monitor is as follows:
7. Remove the probe from the tester and replace it in the probe well.
8. Disconnect the tester.
Model Temperature
9600 106.0 °F ±0.4 °F (41.1 °C ±0.2 °C)
9600 Plus 105.8 °F ±0.4 °F (41.0 °C ±0.2 °C)