32 Functional verification Welch Allyn VSM 300 Series
Here is an example of verification results for a typical monitor.
Test Result
Pass
(or N/A)
Fail
Battery charging charge symbol lit
Power-on self-test normal
Beeper audible
Monitor-off current (mA): 0
Init/idle current (mA): 467
Baseline current (mA): 199
Battery voltage Monitor (V): 5.94 DMM (V): 5.95
NIBP characterization test C03
NIBP verification
Pressure leakage Before (mmHg): 100 After (mmHg): 99
Pressure accuracy 80 mmHg SYS: 73 Manometer: 72
150 mmHg SYS: 142 Manometer:140
300 mmHg SYS: 290 Manometer: 286
Valve/pump current Valve/Pump (mA)
Baseline (mA)
Difference (mA)
560
199
361
Inflation Elapsed time (sec): 6
Deflation Manometer reading (mmHg): 0 (3 sec)
Max allowable cuff pressure
Adult (280 [37.3]) 280 mmHg kPa
Pediatric (200 [26.7]) 200 mmHg kPa
Neonate (141 [18.8]) 141 mmHg kPa
Overpressure cut-off (Pass-Fail)
Printer current Peak (mA): 940
SpO
2
Saturation (%): 90 Pulse (bpm): 200
Temperature Fahrenheit: 105.9 Celsius:
Nurse call relay pins 1-2 Alarm () >1M No Alarm () <1
pins 2-3 Alarm () <1 No Alarm () >1M
Battery charge time 12.5 h
Battery discharge time 19 h
Battery voltage 6.40 volts
Monitor Serial #: JA036582
Tested by: RJ
Test date: 8/30/2007