Electromagnetic emissions
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
WARNING This equipment/system is intended for
use by healthcare professionals only. This
equipment/ system may cause radio interference or
may disrupt the operation of nearby equipment
a
. It
may be necessary to take mitigation measures, such
as re-orienting or relocating the monitor or shielding
the location.
Complies
a
The monitor contains a 5-GHz orthogonal frequency-division multiplexing transmitter or a 2.4-GHz frequency hopping
spread-spectrum transmitter for the purpose of wireless communication. The radio is operated according to the
requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/EC). The transmitter is
excluded from the EMC requirements of 60601-1-2, but should be considered when addressing possible interference
issues between this and other devices.
Electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
monitor should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the monitor requires continued operation
during power mains interruptions, it is
recommended that the monitor be powered
from an uninterruptible power supply or a
battery.
Electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
monitor should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the monitor, including cables, than the
recommended separation distance calculated
148 Guidance and manufacturer's declaration Connex
®
Spot Monitor

Table of Contents

Other manuals for Welch Allyn Connex Spot Monitor

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
Battery LifeUp to 8 hours
Display7 inch color touchscreen
ConnectivityBluetooth, USB
Power SourceAC power, Rechargeable battery
MeasurementsBlood pressure, temperature, SpO2, pulse rate
SpO2 TechnologyMasimo or Nellcor
Optional FeaturesPrinter
Temperature TechnologySureTemp® Plus or Braun Thermoscan® Pro 6000