Message Possible cause Suggested action
Alarm
priority
Invalid SpO2 interval
parameter during
interval capture.
An invalid interval parameter was
detected.
If SpO2 intervals are enabled and SpO2 clip
was removed, either stop the intervals or
reattach the SpO2 clip. Information status
message; press OK button to dismiss.
Information
Radio messages
Message Possible cause Suggested action
Alarm
priority
Radio not functional.
350001
Deserialization failure. There is a
software communication issue
between the host and the radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
350002
Permissions. There is an internal
software error on the radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
350003
Unsupported operating system.
There is an internal software error
on the radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
350004
Unknown. There is an internal
software error on the radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
350006
Invalid authentication. There is an
internal software error on the
radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
350008
Unknown SDC Error. There is an
internal software error on the
radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
350009
Invalid SDC configuration. There is
an internal software error on the
radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Radio not functional.
35000a
Invalid SDC profile. There is an
internal software error on the
radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
Invalid radio
configuration.
Reconfigure and try
again. 35000b
Invalid SDC WEP type. There is an
internal software error on the
monitor. Attempt to configure
settings that don't apply in current
authentication mode on radio
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
Very low
Invalid radio
configuration.
Reconfigure and try
again. 35000c
Invalid SDC EAP type. There is an
internal software error on the
monitor: attempt to configure
settings that don't apply in current
authentication mode on radio
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
Very low
Invalid radio
configuration.
Invalid SDC parameter. The Laird
SDK rejects a parameter being
configured.
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
Very low
Directions for use Troubleshooting 117

Table of Contents

Other manuals for Welch Allyn Connex Spot Monitor

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
Battery LifeUp to 8 hours
Display7 inch color touchscreen
ConnectivityBluetooth, USB
Power SourceAC power, Rechargeable battery
MeasurementsBlood pressure, temperature, SpO2, pulse rate
SpO2 TechnologyMasimo or Nellcor
Optional FeaturesPrinter
Temperature TechnologySureTemp® Plus or Braun Thermoscan® Pro 6000

Related product manuals