6 Setup Welch Allyn Connex® Integrated Wall System
Mount the wall system
1. On the selected wall, find and mark the studs, and choose the system height and
corresponding height for the mounting rail bracket.
Recommendation: Place the mounting rail bracket 63 inches (1.6 m) from the floor,
which places screen center height at approximately 63 inches (1.6 m) from the floor.
Caution This drawing shows the physical relationships of the mounting
brackets to each other and to the wall system after you complete the mounting
instructions. Do not place the wall system on the wall until you have completed
all preliminary steps.