7. Faults
7. Faults
Personnel: Skilled personnel
Protective equipment: Protective gloves, safety goggles
Tools: SW 27 open-ended spanner or torque wrench
WARNING!
Physical injuries and damage to property and the environment
caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable
or toxic substances) or harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), there is a danger of physical injuries and damage to property
and the environment.
Should a failure occur, hazardous media under high pressure or vacuum
may be present at the instrument.
▶
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
▶
Wear the requisite protective equipment.
If faults cannot be eliminated by means of the listed measures, the
instrument must be taken out of operation immediately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter
9.2 “Return”.
For contact details, see chapter 1 “General information” or the back page
of the operating instructions.
Display Causes Measures
Low battery voltage, functioning
is only guaranteed for a short
period of time
Insert new batteries, see chapter
8.2.2 “Battery replacement”
OL
-OL
Reading is significantly above
or below the measuring range
≥ 10 % FS
Check: Is the pressure within the
permissible measuring range of the
sensor?
■
If the pressure is outside the permissible
range, remove the pressure or vacuum
source immediately from the CPG1500
in order to avoid damaging the internal
sensor.
■
If the pressure is within the permissible
range, contact the manufacturer.