58 WIKA operating instructions, model CPG1500
EN
14161966.06 03/2024 EN/DE
8.2.2 Battery replacement
Wear protective gloves!
Use protective gloves when damaged or leaking/leaked batteries must be
removed.
1. Switch off the instrument and lay it face down.
2. Loosen the three screws of the battery compartment, see Fig. 2 “Position of the
battery compartment”.
3. Remove the battery cover.
4. Remove exhausted batteries.
5. Insert the three AA-sized batteries in correct polarity.
Make sure that the polarity, (+) or (-), on the batteries matches the markings in the
battery compartment.
6. Put the battery cover on and tighten it with the screw.
Tighten the upper screw first.
Maximum tightening torque of the screws is < 0.4 Nm
Fig. 2 - Position of the battery compartment
8. Maintenance, cleaning and calibration