TSW200E1
BER: rate, in exponential notation, between the number of errored bits and the total
number of bits received.
INJEC. ERR.: Displays the number of transmission errors inserted by means of the
ERROR key.
RX RATE: on-line frequencimeter, which estimates the received data rate.
TIME: shows the total test elapsed time. Format: hours:minutes:seconds.
From this first basic errors screen press the key:
<F2> (MENU) to go back to MENU screen
<F4> (NEXT) to access the second basic errors screen
The second basic errors screen is shown below:
F1 F2 F3 F4
B A S I C E R R O R S
BLK ERROR = 00000
BLER = 0.00E-00
BLOCKS = 6.80E+04
BACK MENU NEXT
From the second basic errors screen, it is possible to press the key:
<F1> (BACK) to go back to the previous screen
<F2> (MENU) to go back to MENU screen
<F4> (NEXT) to go the third basic errors screen
The counters displayed in the second basic errors screen are the following:
BLK ERROR: indicates the total number of block errors that happen when the test is
running. Any number of errors bits that are flagged in a 1024-bit block is marked as a
single error block.
BLER: indicates the rate, in exponential notation, between the number of error blocks
and the total number of received blocks.
BLOCKS: exponential counter displaying the total number of 1024-bit blocks received
when the test is running.
From the third basic errors screen displayed below, it is possible to press:
<F1> (BACK) to go back to the previous screen
<F2> (MENU) to go back to MENU screen
<F4> (NEXT) to go the fourth basic errors screen
Wise Telecommunications Industry 38