EasyManua.ls Logo

XS Instruments COND 7 Plus - Data Sheet

Default Icon
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
7+Series Version 2.0
Data sheet
Specifications
Models
pH
0.00 ~ 14.00pH
pH 7+
PC 7+
0.1/0.01 pH
±0.01 pH ±1digit
0 ~ 100°Cmanual or automatic
1-3 point
USA: 1,68 - 4,00 - 7,00 - 10,01pH
NIST: 1,68 - 4,01 - 6,86 - 9,18pH
2 value CUSTOMER
mV
±1,000mV
1mV
±0.1% FS ±1digit
Conductivity
Conductivity: 0~200 mS/cm(divided into four ranges):
0~199.9μS/cm ; (200~1999) μS/cm ;
2.00~19.99mS/cm; 20.0~199.9mS/cm
C0ND 7+
PC 7+
0.1/1μS/cm 0.01/0.1 mS/cm
±2.0% FS
0 ~ 80manual or automatic
0.1 / 1 / 10 cm
-1
1-4 point
84 μS/cm, 1413 μS/cm, 12.88, 111,9 mS/cm
1 customer value
TDS
0~100 g/L
C0ND 7+
PC 7+
1% f.s.
0.4 ~ 1.0
Temperature
0~100°C
pH 7+
C0ND 7+
PC 7+
0.1
±0.5±1digit
Reading stability criteria
Low:1.2mV/10 sec., Medium:0.6mV/10 sec., High:0.3mV/10
seconds
pH 7+
PC 7+
Power
AA batteries × 3 (1.5V× 3)
IP rating
IP57
Dimension & Weight
Meter: (86×196×33 )mm / 300g

Table of Contents