48
(b) on silicone plug (8S) above the capillary tube (22) for suction operation (see Fig. 9)
Fig. 9 Suction mode liquid handling with manual pump VZ 6554
The suction operation with manual pump VZ 6554 (optional accessory, not included in the scope of
delivery), is recommended especially for aggressive liquids which must never enter the thermostat bath
inadvertently. Furthermore, this operating mode is recommended for foaming samples.
3.9 Start measurement
Remove the adapter (28) of the manual pump.
Considering that the measurement liquid will immediately start flowing down, the <START>
key (16) must be pressed immediately!
16
8S