Sepalizing in indicator,Being perfect in indicator
Shanghai Yaohua Weighing System Co.,Ltd.
10
strongly recommended
8
Press[0][1][
2],Press[Qua
ntity]for
confirm
Display
[FLT ]
[012 ]
[ ]
Example:012
9
Press[6][0][
0][0]
t
hen
press[Quanti
ty]for
confirm
Display
[FULL ]
[006000]
[******]
Input full range:
6000
10
After
stability,pr
ess[Quantity
]for confirm
[ noLoAd ]
[SurE ]
[ ******]
Zero- point calibration
11
After 5~10
seconds
stability
press[6][0][
0][0]
t
hen
press[Quanti
ty]for
confirm
[AdLoAd1 ]
[006000 ]
[****** ]
Linearity calibration Put
ertain load,the more close to
he full range,the better it
s.After 5~10 seconds
tability,input the actul
value of the loaded weight
Example6000
12
Calibration
over,return
to the
weighing
status
[ ****** ]
[ ****** ]
[ ****** ]
Note 2-3
Note 2-1,Pn parameter setting:
W 0 1 2 3
Zero track speed 0.4second 0.3second 0.2second 0.1second
X 0 1 2 3 4 5 6 7 8 9
Zero track
range
No
track
0.5
e
1.0
e
1.5
e
2.0
e
2.5
e
3.0
e
3.5
e
4.0
e
4.5
e
Y 1 2 3 4 5
manual zero setting
range
2F.S 4%F.S 10%F.S 20%F.S 100%F.S
Z 1 2 3 4 5
auto zero setting
range
2%F.S 4%F.S 10%F.S 20%F.S 100%F.S