Sepalizing in indicator,Being perfect in indicator
Shanghai Yaohua Weighing System Co.,Ltd.
9
Calibration
1. Connect load cell properly,then turn on the indicator,it will
perform self-checking,then enter into the weighing
statement.(15~30 minutes pre-heat is necessary to assure
excellent performance)
2. Put the jumper inside the indicator at ”On” location(the default
location is just “On”),then calibrate as shown table:
Step
Operation
Display
Notes
1
[Quantity] +
[.Value]
[ PASS ]
[000000]
[ ]
E
nter the calibration,notes for
inputting the password
2
Input the
password by
keypad,,Pre
ss[Quantity
]for
confirm
[ PASS ]
[ 319015]
[ ]
Password319015
3
Press[2],Pre
ss[Quantity]
for confirm
[ E ]
[ 002 ]
[ ]
Enter division optional:
1/2/5/10/20/50/100/200,if the
o
riginal divison is accepted,then
press[Quantity]for confirm
Example: 2
4
Press[2],
press[Quanti
ty]for
confirm
[ POS ]
[ 2 ]
[ ]
Enter decimal number (0~4)
Example: 2
5
[2Ero ]
[WXYZ ]
[ ]
Set the Zero parameterDetail in
Note 2-1
WZero track speed XZero track
range
Ymanual zero setting range Z
auto zero setting range
6
Press[1][2][
4],Press[Qua
ntity]for
confirm
[2Ero ]
[0124 ]
[ ]
Example:0124
7
Display:
[FLT ]
[XYZ ]
[ ]
Set the digital filter intensity
Detail in Note 2-2
X anti-vibration select Y stable
range Z filter insensity the
high value it is,t
h
e more filter
i
nsensity and the lower response
i
t is.The parameter 012 is