Errors Information
1. For details of the error information,pls refer to the following
table:
Error Notes Solution
[Err 01]
[ ******]
[ ******]
The analog signal is
seriously negative
C
heck the connection of the
load cell and indicator
[Err 02]
[ ******]
[ ******]
The analog signal is
too high or error
from load cell
C
heck the connection of the
load cell and indicator
[Err 03]
[ ******]
[ ******]
The weight is
overloaded
1
I
f the indicator hasn’t
b
een calibrated,calibrate
it.
2Decrease the weight so
t
hat it is within the full
range
3Re-set the full range
[Err 10]
[ ******]
[ ******]
The unit weight
isn’t enough
1Re-
s
elect the load cell
with high precision
2Increse the samples
[Err 13]
[ ******]
[ ******]
Parameter is
wrongly set
Re-set the parameter
[Err 11]
[ ******]
[ ******]
The loaded analog
signal is too weak
while calibration
D
ecrease the full range or
s
elect load cell with
more range sensitivity
[Err 15]
[ ******]
[ ******]
The full range too
small
F
ull range should be more
than 100
[Err 23]
[ ******]
[ ******]
Memory defective
C
heck the memory on the
m
ain board or the circuit
t
o see whether therer is
s
horted or
opened,otherwise,change
the memroy