7.3.2 Active Power Accuracy
Item Specifications
Accuracy Requirements
Same as the conditions for voltage and current.
Power factor: 1
Accuracy (at 12 months)
(The accuracy shown below is the sum of reading and range errors.)
* f in the read error equation is the input signal frequency in kHz.
WT310/WT332/WT333
WT310HC (Current EXT sensor input)
WT310HC (Direct current input)
DC ±(0.1% of reading + 0.2% of range) ±(0.3% of reading + 0.2% of range)
0.5Hz≤
f<45Hz ±(0.3% of reading + 0.2% of range) ±(0.3% of reading + 0.2% of range)
45Hz≤
f≤66Hz ±(0.1% of reading + 0.1 % of range) ±(0.1% of reading + 0.1 % of range)
66Hz<
f≤1kHz ±(0.2% of reading + 0.2 % of range) ±(0.2% of reading + 0.2 % of range)
1kHz<
f≤10kHz
±(0.1% of reading + 0.3 % of range)
±[{0.067×(f–1)}% of reading]
±((0.13×f)% of reading + 0.3 % of range)
10kHz<
f≤20kHz ±((0.13×f)% of reading + 0.5 % of range)
10kHz<
f≤100kHz
±(0.5 % of reading + 0.5 % of range)
±[{0.09×(f–10)}% of reading]
Influence of temperature changes after zero-level compensation or range change
Add the product of the voltage influence and the current influence listed below to the DC power
accuracies.
DC voltage accuracy: 0.02% of range/°C
DC current accuracies
WT310(5mA/10mA/20mA/50mA/100mA/200mAranges):5μA/°C
WT310(0.5A/1A/2A/5A/10A/20Aranges)andWT332/WT333directcurrentinput:500μA/°C
WT310HC direct current input: 1mA/°C
External current sensor input (/EX1): 1mV/°C
Externalcurrentsensorinput(/EX2):50μV/°C
Influence of self-generated heat caused by voltage input
Add 0.0000001 ×U
2
% of reading to the AC power accuracies.
Add 0.0000001 ×U
2
% of reading + 0.0000001 ×U
2
% of range to the DC power accuracies. U is the
voltage reading (V).
Influence of self-generated heat caused by voltage input lasts until falling the temperature of the input
resistor even if voltage input decreases.
Influence of self-generated heat caused by current input
WT310:
Add 0.00013 × I
2
% of reading to the AC power accuracies.
Add 0.00013 × I
2
% of reading + 0.004 × I
2
mA (0.5 A, 1 A, 2 A, 5 A, 10 A, and 20 A ranges) or 0.00013
× I
2
% of reading + 0.00004 × I
2
mA (5 mA, 10 mA, 20 mA, 50 mA, 100 mA, and 200 mA ranges) to
the DC power accuracies.
I is the current reading (A).
WT310HC:
Add 0.00006 × I
2
% of reading to the AC power accuracies.
Add 0.00006 × I
2
% of reading + 0.001 × I
2
mA to the DC power accuracies.
I is the current reading (A).
WT332/WT333:
Add 0.00013 × I
2
% of reading to the AC power accuracies.
Add 0.00013 × I
2
% of reading + 0.002 × I
2
mA to the DC power accuracies.
I is the current reading (A).
Influence of self-generated heat caused by current input lasts until falling the temperature of the shunt
resistor even if current input decreases.
Accuracy changes caused by data update interval
When the data update interval is 100 ms, add 0.05% of reading to the 0.5 Hz to 1 kHz accuracies.
Guaranteed accuracy ranges for frequency, voltage, and current
All accuracy figures for 0.5 to 10 Hz are reference values.
The power accuracy figures for DC, 10 Hz to 45 Hz, and 400 Hz to 100 kHz when the current exceeds
20 A are reference values.
7.3 Accuracy

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT310 and is the answer not in the manual?

YOKOGAWA WT310 Specifications

General IconGeneral
ModelWT310
CategoryDigital Power Meter
Current Measurement Range0.5 mA to 20 A
WeightApprox. 3.5 kg
Voltage Measurement Range15 V to 600 V
Frequency Range100 kHz
Basic Accuracy0.1%
Power Accuracy0.1%
DisplayLCD
InterfacesUSB, RS-232, GP-IB
Input Elements1
Dimensions213 mm x 88 mm x 346 mm