0 ~ 20mA (installed I4
module in MIO)
Four formats (0, 0.0, 0.00, 0.000) are selectable
Note 1: For thermocouples or RTD input, only 0 or 0.0 is selectable, and the internal
resolution is 0.1. When S type thermocouple is used, dPt is recommended to be 0. If
Inp= 17,18 or 22, resolution will support display 0.0 or 0.00
Define scale low limit of input. It is also the low limit of transmitter output (CtrL=POP
or SOP) and light bar display.
Define scale high limit of input. It is also the high limit of retransmission output
(CtrL=POP or SOP) and light bar display.
Scb is used to shift input to compensate the error caused by transducer, input signal,
or auto cold junction compensation of thermocouple.
PV after compensation=PV before compensation + Scb
It is generally set to 0. The incorrect setting will cause measurement inaccurate.