9
-
FEATURES
15. Safety System:
Over Current or Connection Loosen (Loss Power): Alarming and Shut Down
Compressor Over Hot: Unit Shut Down
Pressure Failure: Alarming and Shut Down
Compressor Failure: Alarming and Shut Down
Low Oxygen Concentration
16. Normal Operating Condition (with Oxygen Concentration Status Indicator):
Temperature range: 5℃40℃
Relative humidity: 15%90%
Atmosphere pressure: 860hPa1060hPa
Note: When operating condition over the rated ranges of ambient temperature, humidity
and atmosphere pressure, the performance of the oxygen may be dropped.
17. Oxygen Output Temperature: ≤ 46℃
Applied part TemperatureNasal Cannula: ≤ 41℃ in normal condition
Applied part TemperatureNasal Cannula: ≤ 43℃ in single fault condition
18. Cannula length DO NOT more than 15.2m and no twist.
19. Storage and Transportation Condition:
Temperature range: -20℃60℃
Relative humidity: 10%93% Non-condensing.
Atmosphere pressure: 700hPa1060hPa
The device should be stored with no strong sunlight, no corrosive
gas and well ventilated indoor area. The device must be
transported and used in the vertical position only.
When the oxygen concentrator is stored outside the temperature
range specified under normal operating condition, please put the
device in normal operating condition more than 4 hours before
operating.
20. The maximum oxygen flow for the accessories is not greater than 10L/min. And
the maximum pressure for the accessories is not greater than 200 kPa.