17
-
MAINTENANCE
NOTE: Power should be disconnected before beginning preventive maintenance on the
concentrator. The concentrator is specifically designed to minimize routine preventive
maintenance at intervals of once per year. In places with high dust or soot levels,
maintenance may need to be performed more often. The following must be performed at a
minimum of one year in service to assure years of additional reliability.
. CLEAN CABINET
CUT OFF the power supply first to avoid electrical shock.
DO NOT remove device cabinet.
Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge at
least one time a month. Do Not drop any liquid into the seam of the device.
. CLEAN OR REPLACE FILTER
Please clean or replace the filters in time, it’s very important to protect compressor and
to extend the device life.
1. Clean the I-stage air inlet once a month. ( Figure 6 )
2. The II-stage filter mesh cleans once half a month.
Replace or clean the filter immediately if the mesh turns black. (Figure7)
1) Clean the filter with a soft cleaner or wash in warm soapy water and rinse thoroughly.
2) DRY the filter thoroughly before reinstallation.
3) The filter must clean or replace after operating 100 hours operating at least or as
necessary.