9
8
22.Transportation and storage conditions:
●Temperature range: -20℃~+55℃(-4℉~+131℉),
21.Suggestion: The length of nasal oxygen cannula should be
less than 15.2 meters and should not be pinched.
20.Temperature at oxygen outlet: ≤46℃
●Atmosphere pressure: 700hPa~1060hPa (10.15psi ~ 15.37psi).
NOTE: When storage temperature is below 5℃, please keep
the equipment in normal working condition for at least 4
hours before using.
●Relative humidity: 10%~93%, no condensation,
NOTE: Oxygen concentrator should be stored with out
intense sunlight, no corrosive gas and ina well ventilated
room, avoid jolt and inverted recumbent transport.
FEATURES
NOTE: Unless the oxygen concentrator is to be used
immediately, otherwise it should be stored in the containers
until it's being used again.
Check for any obvious damage to the carton or its contents.
If damage is evident, please notify the carrier or local dealer.
Remove all loose packing from the carton.
Carefully remove all the components from the carton.
Examine exterior of the oxygen concentrator for nicks,
dents, scratches or other damages. Inspect all components.
DO NOT place other objects on top of the repackaged
concentrator.
Store the repackaged oxygen concentrator in a dry area.
Ⅰ. UNPACKING
Ⅲ. STORAGE
Ⅱ. INSPECTION
HANDLING