Workflow.
Manual cleaning
Place the components in the cleaning solution for the duration of the specified immersion time,
so that the components are sufficiently covered. Ensure that the components do not touch each
other and that no air remains in cavities.
Support the cleaning process by careful brushing with a soft brush and ultrasonic treatment.
(duration in addition to brushing at least as long as the required immersion time).
If applicable: Clean all cavities of the components at the beginning and after the immersion
time has elapsed, flush at least five times using a disposable syringe (minimum volume 5 ml).
Brush out openings with a conical interdental brush.
Remove the components from the cleaning solution and rinse them intensively at least three
times (for at least one minute) rinse with water.
If applicable: Rinse all cavities of the components at the beginning and at the end of the immersion
time, flush at least five times using a disposable syringe (minimum volume 5 ml).
Check components, see chapter: 8.4. Inspection, p.39.
Disinfection
Immerse the components in the disinfectant solution for the specified immersion time so that
the components are sufficiently covered. Ensure that the components do not touch each other
and that no air remains in cavities.
If applicable: All cavities of the components must be rinsed at least five times with a disposable
disinfectant at the beginning and at the end of the immersion time, flush at least five times using
a disposable syringe (minimum volume 5 ml).
Remove the components from the cleaning solution and rinse intensively at least five times
(for at least one minute) with water.
If applicable: flush all cavities of the components at the beginning and at the end of the immersion
time, flush at least five times using a disposable syringe (minimum volume 5 ml).
Dry and pack the components immediately after removal (see chapter: 8.5. Packaging, p.40).