EasyManua.ls Logo

Zodiac TRi PRO - Page 36

Zodiac TRi PRO
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
3. Leave the ACL sensor immersed in the buffer solution for approximately 5 minutes in order to ensure
a reliable measurement.
4. From the default display on the LCD screen, press the (MENU) button.
5. Use the (UP)/(DOWN) buttons to displaypH/ACL SETUP’, then press (SELECT).
6. Use the (UP)/(DOWN) buttons to display CALIBRATE ACL’, then press (SELECT).
7. Displayed ACL value on the LCD screen (‘CURRENT ACL’) could have 3 different values : HIGH’,
LOWorOK’. Press (SELECT) to start the calibration process (see pictures below).
8. Once the calibration is completed, ACL value should beOK(‘CURRENT ACL’). If the displayed value
is notOK’, repeat steps 1 to 7.
21. 9. Press (MENU) button twice or wait 30 seconds, the TRi power pack will return to the default
display on the LCD screen.
IMPORTANT NOTES
:
- In order to ensure that the Zodiac TRi remains fully efficient, it is recommended to perform a
calibration at least once every two months during the season of pool use.

Table of Contents

Related product manuals