Preventive Maintenance Test
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 1–5
Test Procedure
Test Action Expected Result(s) Pass Fail
2.1 Turn on the AED Plus unit.
! UNIT OK
! ADULT PADS
! DON’T TOUCH PATIENT,
ANALYZING
! TREATMENT ADVISED
(or SHOCK ADVISED)
! DON’T TOUCH PATIENT.
! PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
!!
2.2 Press the Treatment button
(first shock).
• Analyzer indicates a defibrillator
discharge at the configured
first-shock energy level (120, 150
or 200 joules ±15%).
! TREATMENT DELIVERED
(or SHOCK DELIVERED)
• An updated shock count appears
on the display (SHOCKS: 1).
! DON’T TOUCH PATIENT,
ANALYZING
! TREATMENT ADVISED
(or SHOCK ADVISED)
! DON’T TOUCH PATIENT.
! PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
!!