CHAPTER 1MAINTENANCE TESTS
1–6 www.zoll.com 9650-0308-01 Rev. A
2.3 Press the Treatment button
(second shock).
Analyzer indicates a defibrillator
discharge at the configured
second-shock energy level (120,
150 or 200 joules ±15%).
! TREATMENT DELIVERED
(or SHOCK DELIVERED)
An updated shock count appears
on the display (SHOCKS: 2).
! DON’T TOUCH PATIENT,
ANALYZING
! TREATMENT ADVISED
(or SHOCK ADVISED)
! DON’T TOUCH PATIENT.
! PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
!!
2.4 Press the Treatment button
(third shock).
Analyzer indicates a defibrillator
discharge at the configured
third-shock energy level (120, 150
or 200 joules ±15%).
! TREATMENT DELIVERED
(or SHOCK DELIVERED)
An updated shock count appears
on the display (SHOCKS: 3).
These three prompts are optional:
OPEN AIRWAY
CHECK BREATHING
GIVE TWO BREATHS
! CHECK CIRCULATION
(or CHECK PULSE)
! IF NO CIRCULATION START CPR
(or IF NO PULSE START CPR)
!!
Test Action Expected Result(s) Pass Fail