10 11
1
b) Results Detailpage
3M™ Attest™ Auto-reader 490H Results Detail
Status
History
Print
Result
Start Time
Well# and Configuration
490H Serial Number
Control?
BI Lot#
Load#
Sterilizer#
Technician
Chemical Integrator?
Implant?
Cycle Parameters
Notes
no
2018-12DA
1
6
sdf
n/a
n/a
n/a
Daily control BI
–
2017-05-11 08:05:42
#2 1295
200299
This page provides an individual record for an incubated BI. For complete documentation,
the user may print a hard copy and/or save the record as an electronicfile.
Note 1: The top four rows (highlighted in green) are populated automatically by the
3M™ Attest™ Auto‑reader 490H and cannot be changed by the user. The remaining rows
are completed by the user to satisfy the documentation requirements of theirfacility.
Note 2: The Results Detail page refreshes less frequently that the Status page. Until the BI
result is available, the Results Detail page displays remaining minutes of incubation in the
resultsfield.
c) Historypage
Status
History
3M™ Attest™ Auto-reader 490H
Well# and Configuration
#3 1295
#2 1295
#1 1295
#10 1295
#3 1295
#2 1295
#1 1295
#4 1295
#3 1295
#2 1295
Start Time
2017-05-11 13:51:49
2017-05-11 08:05:42
2017-05-11 08:03:56
2017-05-07 11:55:16
2017-05-07 11:06:44
2017-05-07 08:03:19
2017-05-07 08:02:43
2017-05-04 08:57:51
2017-05-04 08:57:12
2017-05-04 08:01:47
Result
---
–
+
–
–
–
+
–
–
–
ID
1
2
3
4
5
6
7
8
9
10
The 3M™ Attest™ Auto‑reader 490H stores results for the last 100incubated BIs on
embedded software. The 3M™ Attest™ Auto‑reader 490H Web App History page provides
a summary of these last 100results.
d) Configurationpage
This page allows the configuration of the network and language settings for the
3M™ Attest™ Auto‑reader 490H from the 3M™ Attest™ Auto‑reader 490H Web App.
Documenting BI Results using the 3M™ Attest™ Auto‑reader 490H
Web App
Status
History
3M™ Attest™ Auto-reader 490H
AT TEST 1295 AT TEST 1295
Auto-reader 490H
1 2 3 4 5
+24–
678910
SN
200299
2017- 0 5 -11 17:13
1. Open the Results Detailpage
After inserting a new 3M™ Attest™ Rapid Readout BI in the 3M™ Attest™
Auto‑reader 490H, it is a best practice to open the 3M™ Attest™ Auto‑reader 490H
Web App Results Detail page for that BI by clicking on the appropriate well number
from the Statuspage.
2. Enter the specific information for that BI (e.g., BI Lot #, sterilizer #, load #, technician
name, etc.) in the provided editable fields. Use the Notes section to capture
information not prescribed in the provided row titles; for example, you may wish to use
this field to record loaditems.
3M™ Attest™ Auto-reader 490H Results Detail
Status
History
Print
Result
Start Time
Well# and Configuration
490H Serial Number
Control?
BI Lot#
Load#
Sterilizer#
Technician
Chemical Integrator?
Implant?
Cycle Parameters
Notes
no
2018-12DA
1
6
sdf
n/a
n/a
n/a
Daily control BI
–
2017-05-11 08:05:42
#2 1295
200299
3. The entered information is saved automatically to temporary web browser storage on
the user’s PC every 60seconds, or by leaving the Results Detailspage.
4. Print/File. Once the BI result is available, print or file the Results Detail page to create
a record. The page may be physically printed and/or saved as an electronicfile.
Important! Data is only stored for the last 100incubatedBIs.
Embedded software in the 3M™ Attest™ Auto‑reader 490H saves the data from the
auto‑populated (i.e., green highlighted) fields. The application relies on the temporary
storage in the user’s web browser to save the data in the user‑populated fields. In
both cases, data is only stored for the last 100records, i.e., the 101st incubated BI will
overwrite the information saved for the 1st incubated BI. To retain the Results Detail
information for future reference it is best practice to print the page or save to an electronic
file as soon as the BI result isavailable.