EasyManua.ls Logo

3M Attest Auto-reader 490H - Troubleshooting the 3 M Attest Auto-Reader 490 H Webapp; Repair and Replacement

3M Attest Auto-reader 490H
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
1
TROUBLESHOOTING THE 3M ATTEST AUTOREADER 490H WEB APP
Issue Problem Solution
Unable to access 3M™ Attest™
Autoreader 490H Web App on
networked computer
3M™ Attest™ Autoreader 490H and/or
computer not connected tonetwork.
Autonegotiation issue between the
3M™ Attest™ Autoreader 490H and facility
networkswitch.
Ensure 3M™ Attest™ Autoreader 490H Ethernet cable is connected to a live jack and
that connection to 3M™ Attest™ Autoreader 490H issecure.
Ensure computer is connected tonetwork.
If failure is with connection attempted using 3M™ Attest™ Autoreader 490H serial
number URL, try connecting using IP addressURL.
Switch port setting must be set at 10Mbps andhalfduplex.
Contact facility IT or 3M if problem persists.
Unable to access 3M™ Attest™
Autoreader 490H Web App on
mobile device
3M™ Attest™ Autoreader 490H not connected
tonetwork.
For security reasons, facility firewall may prevent
mobile devices not connected to the facility
network from connecting to secure IP addresses.
Ensure 3M™ Attest™ Autoreader 490H Ethernet cable is connected to a live jack and
that connection to 3M™ Attest™ Autoreader 490H issecure.
Contact facility IT for their recommendation on connecting to the secure network.
Solutions may include enabling WiFi on the mobile device and signing into the facility
network or connecting to the facility virtual private network (VPN) on the mobile device.
Data entered by user on Results
Detail page disappears
User entered data is written to temporary
webbrowser storage. Browser may have
cleareddata.
Set web browser to accept (allow or enable)cookies.
As a best practice, print or save Results Detail record as soon as BI result is final.
BI result displayed on
3M™ Attest™ Autoreader 490H
LCD display does not appear on
Results Detail page
The Results Detail page has not refreshed yet.
Use your browser to refresh the Results Detail page, or wait up to 60seconds for the
page to automatically refresh.
Results Detail page will not open
from the Status page
The Results Detail page only opens from the
Status page for occupied 3M™ Attest™
Autoreader 490H wells.
Do not attempt to navigate to the Results Detail page from unoccupied wells on the
Status page.
REPAIR ANDREPLACEMENT
In the U.S.A., 3M™ Attest™ Autoreaders 490H must be repairedat:
3M Health Care ServiceCenter
Suite 200, Bldg.502
3350Granada AvenueNorth
Oakdale, MN55128
18002926298
Fax:18007708016
Outside of the U.S.A., contact your local 3M Subsidiary.
If you have questions, call our Customer Service number:18002926298.
Made in the U.S.A. of globally sourced materials by 3M Health Care.
3M Health Care
2510Conway Ave
St. Paul, MN55144
18002283957
www.3M.com/healthcare
© 2017, 3M. All rightsreserved.
3M and Attest are trademarks of3M.
Used under license inCanada.
Issue Date: XXXX-XX
3M Deutschland GmbH
Health Care Business
Carl-Schurz-Str. 1
41453 Neuss
Germany

Related product manuals