INSTRUCTIONS FOR USE WELLS, BUTTONS, & DISPLAYINTERPRETATION
Please familiarize yourself with the face of the 3M™ Attest™ Autoreader490H:
Wells
1. Incubation/Reader Wells The 3M™ Attest™ Autoreader 490H has ten
incubationwells.
Buttons
2. Information / Up ArrowButton
3. Temperature / Down ArrowButton
4. Buzzer Alarm OFF / EnterButton
InformationDisplay
5. LCD display Display isbacklit
6. LCD display Example of Error Message for well#6
7. LCD display Example of Remaining Incubation Time, in minutes, for well#10
8. LCD display Example of Negative ( ) BI result for well#9
9. LCD display Example of Positive ( + ) BI result for well#8
10. LCD display Alarm EnabledIndicator
ColorCoded ConfigurationIndicator
11. Incubation well configuration sticker is colorcodedpink
INSTRUCTIONS FOR USE
3M
ATTEST AUTOREADER490HCONFIGURATION
Note: Ideally, the 3M™ Attest™ Autoreader 490H should be empty (i.e., have no
incubating BIs) when it isreconfigured.
Press the
and buttons on the front of the 3M™ Attest™ Autoreader 490H
simultaneously for 3seconds to enter configuration mode. Once in configuration mode,
use the Up
and Down buttons to set the 3M™ Attest™ Autoreader 490H
clock and audible alarm as desired. The button
acts as an Enter key and moves the
flashing cursor to the next changeablesetting.
Setting the 3M Attest Autoreader 490HClock
The 3M™ Attest™ Autoreader 490H date and time are displayed in this format:
YYYYMMDD HH:MM, with the time displayed using a 24hour clock. To change the
year, if necessary, push the up or down buttons to set desired year, then press
to
advance to the month, adjust as necessary and press
to advance to the day. Use the
button to move to the time and set the hours andminutes.
Setting the Audible AlarmFunction
Move across the top row of text by pushing the button. The flashing B in Buzzer
indicates this is a changeable setting. It is recommended this be set at Buzzer Enabled
to audibly alert the operator to a positive BI but it may be set to Buzzer Disabled by
using either the up or down button. Note: If Buzzer Disabled is selected, and the change is
accepted, the speaker icon (
) will disappear from the LCD panel and the alarm will be
disabled for all positive 3M™ Attest™ Rapid Readout Biological Indicators, error codes,
and caution code alarmwarnings.
Note: The 3M™ Attest™ Autoreader 490H display backlight flashes whenever an
audible alarm sounds. Disabling the audible alarm does NOT disable the LCD display
backlightflashing.
Press
to select the desired option and advance to the Acceptscreen.
AcceptScreen
Use the up or down buttons to toggle between Yes and No, press to accept changes
(Yes selected) or retain current settings (No selected) and exit the configuration mode. The
LCD panel will return to the standarddisplay
.