3M™ PROTÉGÉ PLUS
PORTABLE MULTI-GAS MONITOR
MAINTENANCE
8006569, Rev. C/June 2019 7-11
Figure 7-9: Remove the Alkaline Battery Cover
3. Remove the old batteries from the battery holder and replace them with new ones. Ensure you
note their polarity (refer to Figure 7-10:).
Figure 7-10: Replacing with the New Alkaline Battery
4. Replace with the new batteries.
NOTE: Ensure the polarity is correct when installing the new battery.
WARNING: AFTER REPLACEMENT, PERFORM A BUMP TEST AND CALIBRATION TO
ENSURE PROPER OPERATION PRIOR TO PLACING BACK INTO SERVICE.
FAILURE TO FOLLOW THIS WARNING COULD LEAD TO SERIOUS
INJURY OR DEATH.
1
2
3
4
M2.5 x 20
Screws
M2.5 x 10
Screws
Alkaline
Battery Pack
Battery
Cover