70-2011-5643-0 Issue Date: 10/2015
3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series
Site Planning and Installation Guide
Electrical Service Requirements
The electrical service requirements for the installed 3M equipment are shown in Table 3. Be sure to
provide sufficient access to the power disconnect and switch. When planning the electrical service,
consider support equipment including: compressors, air dryers, and emission control systems.
Table 3 – Electrical Service Requirements
WARNING
To reduce the risk of shock due to
hazardous voltage, purchaser must provide
a properly grounded outlet (an earth ground) for
installation, per applicable codes and regulations.
WARNING
To reduce the risks associated with fire
and explosion, purchaser must comply with
applicable codes and regulations as they pertain
to locating the equipment and providing services
to the equipment.
Requirement GS5 Sterilizer/Aerator GS8 Sterilizer/Aerator
Voltage 200–240 Vac
Frequency 50/60 Hz
Phase Single
Current 15 amp dedicated circuit
Heat Load 5500 Btu/hr 6150 Btu/hr
Internal Circuit Breaker 7 amp 12 amp
Power Cord 240 volt, 15 amp, NEMA 6-15 plug
2
Marks FCC cULus RCM CE0086 FCC cULus RCM CE
Code and Building
Regulation Compliance
Due to variation in local regulations, the purchaser is responsible for complying
with applicable codes and regulations as they pertain to locating the equipment
and providing services to the equipment.
Seismic Bracing
Seismic Bracing kits for 3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series, Models GS5
and GS8 units are available from 3M. For information on the State of California
pre-approved seismic bracing certificates, contact 3M.
UL listed to U.S. and Canadian Safety Standards
RCM Compliant with all applicable regulatory arrangements (RCM)
FCC (Federal Communication Commission)
Conformity to European Directives
16