21
70-2011-5643-0 Issue Date: 10/2015
3M
Steri-Vac
Sterilizer/Aerator GS Series
Site Planning and Installation Guide
Roof
Copper Line
(Maximum Length =
91.5 m/300 ft)
Cupola
Minimum radius; two times
the tubing diameter
Insulation
Steri-Vac
Sterilizer/Aerator
5 cm
(2 in.)
5 cm
(2 in.)
30.5 cm
(12 in.)
18 ga galvanized pipe
13 cm
(5 in.)
Figure 12. Ethylene Oxide Vertical Vent Line Installation Example