GLUCOFIX
®
TECH - 51415 - 02/20
UK
18
4. MEASURING YOUR BLOOD GLUCOSE LEVELS
NOTE
For accurate test results, apply the drop of blood to the tip of the test strip within 20 seconds
after puncturing.
Do not test blood that runs or spreads out from the puncture site.
Do not smear blood onto the test strip.
Do not forcefully press the test strip into your puncture site.
Do not touch the test strip once the meter has started the countdown.
c. Read your test result. You will hear a beep when the test result appears on the screen.
CAUTION
If “LO” or “HI” appears on the screen:
Repeat the test. If “LO” or “HI” still appears, contact your doctor or healthcare professional.
“LO” appears if your test result is less than 1.1 mmol/L “HI” appears if your test result is more
than 33.3 mmol/L.
The “hypo” or “hyper” icons may appear depending on the threshold limit you have set (§6.4).
If test results do not match how you feel:
Make sure you have performed the test correctly as explained in §4.2. Then, conduct a control
test to check that the system is working properly (§4.3). Repeat the test using a blood sample
taken from a ngertip (do not use an alternative site). If the test result still does not match how
you feel, contact your doctor or healthcare professional.