3. BEFORE TESTING
3.1 Caution when using the Meter
Time and date are pre-set in the meter. Please make sure that time and date settings are correct
before rst use, and adjust them if needed. Always check the settings after changing the battery
(§7.3).
CAUTION
For accurate test results, allow the meter, test strips and control solution to adjust to the
surroundings for 30 minutes before testing your blood glucose levels:
- temperature: 5 to 45 °C (41 to 113 °F);
- humidity: 20 to 90% RH (Relative Humidity).
Do not store or use the meter where:
- there are sharp temperature uctuations;
- humidity is high and causes condensation (bathrooms, drying rooms, kitchen, etc.);
- there is a strong electromagnetic eld (close to a microwave oven, cell phone, etc.).
Keep the meter out of the reach of children. Coin batteries may represent a choking hazard.
Do not use the meter if it has been dropped in a liquid or liquids have entered inside, even if
dried afterwards.
Avoid hand contact with test strip port on the meter. A thermo sensor is housed inside the
meter to minimize any errors.