GLUCOFIX
®
TECH - 51415 - 02/20
UK
40
Calibration and
traceability:
The results are equivalent to the plasma glucose concentration (capillary
plasma referenced). The GLUCOFIX
®
TECH system is calibrated using
capillary plasma values determined with a Yellow Springs 2300 analyser
(YSI). The YSI analyser is calibrated (as a secondary reference measurement
procedure) using a series of YSI standards (primary calibrators) which are
taken from the NIST (National Institute of Standards and Technology, USA)
Temperature
compensation:
Automatic compensation using a built-in thermo sensor
Batteries: Two 3.0 V lithium batteries (CR2032)
Battery life: A minimum of 1000 tests or about 1 year (2-3 tests/day)
Memory: 730 test results with markers, dates and times. When the memory is full, the
new results will replace the oldest ones
Data management: Before/after meals, exercise, check markers available
Alarms: Up to three settable acoustic alarms
Data transfer: Via dedicated data cable, NFC, or a Bluetooth device
Averages: Over 1, 7, 14, 30, 60, 90 days
9. TECHNICAL SPECIFICATIONS