EasyManua.ls Logo

A&D GF-2002A - Percent Mode (%) for Target Weighing

A&D GF-2002A
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GX-A Series/GF-A Series Communication Manual 26 / 41
Example of the
CAL command GX-A SeriesCalibration with built-in weight
For an example of the CAL command for the GF-A series, see the example of the EXC command.
Example of error code output of the R command (re-zero)
PC side
A
CAL command
L CR LF
Balance side
AK CR LF
Before execution
Command received
Executingprocessing
AK CR LF
Command completed
Zero display
Time
C
PC side
C
R command
R CR LF
Balance
AK CR LF
Before execution
Command received
Waiting for re-zero stability
processing, time up due
to instability
E CR LF
Error code output
Zero display
Time
C , E 1 1
Error display
A
CAL command
L CR LF
Or stand by for 5 seconds