EasyManua.ls Logo

A&D GF-2002A - Page 27

A&D GF-2002A
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GX-A Series/GF-A Series Communication Manual 27 / 41
Example of the EXC command Calibration with a separate weight(GF-A Series)
PC side
X
EXC command
C CR LF
Balance side
AK CR LF
Before execution
Command received
Waiting for the zero setting
AK CR LF
Command completed
Zero display
Time
E
R T CR LF P
PRT command
Inputting the zero setting
(processing)
AK CR LF
Command received
Waiting for weight value
input
R T CR LF P
PRT command
AK CR LF
Command received
Inputting the value of the
weight(processing)
AK CR LF
Command completed
Waiting for weight removal
Waiting for re-zero stability
(processing)
AK CR LF
Command completed
Place calibration weight
Removing the weight
A
ction:
A
ction: