GX-A Series/GF-A Series Communication Manual 27 / 41
Example of the EXC command Calibration with a separate weight(GF-A Series)
PC side
X
EXC command
C CR LF
Balance side
AK CR LF
Before execution
Command received
Waiting for the zero setting
AK CR LF
Command completed
Zero display
Time
E
R T CR LF P
PRT command
Inputting the zero setting
(processing)
AK CR LF
Command received
Waiting for weight value
input
R T CR LF P
PRT command
AK CR LF
Command received
Inputting the value of the
weight(processing)
AK CR LF
Command completed
Waiting for weight removal
Waiting for re-zero stability
(processing)
AK CR LF
Command completed
Place calibration weight
Removing the weight