83
GX-2000 GX-3000 GX-4000 GX-6100 GX-6000 GX-8000
Weighing capacity
2100 g 3100 g 4100 g 6100 g 6100 g 8100 g
Maximum display
2100.84 g 3100.84 g 4100.84 g 6100.84 g 6108.4 g 8108.4 g
Minimum weighing value (1 digit)
0.01 g 0.1 g
Repeatability (Standard deviation)
0.01 g 0.1 g
Linearity
±
0.02 g
±
0.03 g
±
0.1 g
Stabilization time (typical at FAST)
Approx. 1 second
Approx.
1.5 s
Approx. 1 second
±2 ppm/°C
±5 ppm/°C
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F )
(When automatic self calibration is not used)
±
0.10 g
±
0.15 g
±
0.5 g
Accuracy right after calibration using
the internal mass. See notes below
The value above is to the weighing capacity.
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate
5 times/second or 10 times/second
Minimum unit mass
0.01 g 0.1 g
Counting
mode
Number of samples
10, 25, 50 or 100 pieces
Minimum 100% reference mass
1.00 g
10.0 g
Percent
mode
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
2000 g
1000 g
3000 g
2000 g
1000 g
4000 g
3000 g
2000 g
4000 g
3000 g
2000 g
6000 g
5000 g
4000 g
3000 g
2000 g
8000 g
7000 g
6000 g
5000 g
4000 g
3000 g
2000 g
Weighing pan
165 x 165 mm
Net weight
Approx. 5.1 kg
External dimensions
210 (W) x 317 (D) x 86 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface (Provided as standard)
RS-232C with Windows Communication Tools WinCT
Notes
Accuracy right after calibration using the internal mass in good ambient conditions (within
the temperature range of 10°C to 30°C (50°F to 86°F) with no abrupt changes in temperature
or humidity, no drafts, no effect by magnetic fields or static electricity).
The internal mass has a mass of approximately 500 grams.
Check the internal mass periodically as described in “19. MAINTENANCE”.